Features of the Servotome Electro Scalpel Range II of Satelec
The SERVOTOME electroscalpel through its elaborate system allows the complete rectification and filtering of the supply voltage.
The resulting high frequency current purity provides a high quality cut-off
Allows pure cutting with fully adjustable coagulation by the operator coagulation and fulguration
The SERVOTOME is equipped with a feedback electronics called FEED BACK that instantly and automatically adjusts the current intensity.
Delivered with case of 10 electrodes that cover any need of use.
Technical Details
Identification
- Part in direct contact with the patient: Electrodes/Electrode insulator
- Parts in indirect contact with the patient: Electrode holder
Power settings
- Incision setting (relative units): 1 - 10
- Coagulation setting (relative units): 1 - 10
Control unit
- Width (in mm): 250
- Height (in mm): 110
- Depth (in mm): 240
- Weight (in g): 1200
- Ingress protection rating: IPX0
Generator
- Type of leakage currents: LF
- Supply voltage: 115 VAC / 230 VAC
- Power supply frequency: 50 Hz / 60 Hz
- Power consumption: 170 VA to 230 VA
- Power output: 30 W
- Output voltage: 650 V PP - P = 10, W = 1
- Characteristic impedance: 600 Ω
- Output impedance range: 100 Ω to 2 kΩ
- Output frequency: 1.2 MHz +/- 0.2 MHz
- High-frequency output: type Floating (isolated from ground)
- Power setting range: Less than 1 W to 30 W
- Operating mode: 5 cycles (10 sec. ON / 30 sec. OFF + 10 min stand-by)
- Electrical rating: I
- Internal fuse not accessible to the user: F1: 5mm - 500 mA T / 250 VAC
- Fuse (mains connector) - 115 V: 5 mm x 20 mm / 2 A T type FST - SCHURTER AG
- Fuse (mains connector) - 230 V: 5 mm x 20 mm / 1.25 A T type FST - SCHURTER AG
Length of cords
- Electrode holder cord (in mm): >2000
- Bracelet cord (in mm): >2000
- Footswitch cord (in mm): >2000
Footswitch
- Width (in mm): 70
- Height (in mm): 30
- Depth (in mm): 95
- Mass (in g): 150
- Ingress protection rating: IPX1
Environmental characteristics
- Operating temperature: +10°C to +30°C
- Storage temperature: 0°C to +50°C
- Operating humidity: 30% to 70%
- Atmospheric storage pressure: 500 hPa to 1060 hPa
Environmental restrictions
- Usage premises: Usable in all medical premises. The medical device must not be used in an operating theatre.
- Use in gas-filled atmosphere: The medical device is not designed for use in a type AP or APG gasfilled atmosphere or in the presence of anaesthetic gases.
Immersion: The medical device must not be immersed.