- Half value layer (HVL): 2,2 mm Al.
- Total filtration at 70 kV: 2,4 mm Al.
- Tube inherent filtration at 70 kV: > 1 mm Al.
- Focal spot size: 0,7 mm.
- X-ray tube tension accuracy: ±10%.
- X-ray tube current accuracy: ±20%.
- Radiation linearity: ±10%.
- X-ray emission time accuracy: ±20 ms 0,020 s ≤ t ≤ 0,320 s, ±5% 0,400 s ≤ t ≤ 3,2 s.
- Reproducibility: 0,05.
- Generator: At constant potential.
- X-ray tube nominal current: 4 mA / 8 mA.
- X-ray tube nominal voltage: 60 / 70 kV.
- Exposure times: 0,020 s ÷ 3,2 s (23 steps).
- Reference current-time product: 0,8 mAs 8 mA 0,1 s, 0,4 mAs 4 mA 0,1 s.
- Intensity of radiation in the air: > 30 μGy/h at 1 m from focal spot.
- Leakage radiation (measured @ 70kV, 8 mA, 3,2 s): < 0,25 mGy/h at 1 m from focal spot
- Operating cycle: 1:32.
- Focal spot: 0.7 mm.